The [c]FP-RTD-122 uses a linearization curve known as the 
Callendar-Van Dusen equation to measure the temperature of 
RTDs. The equation is as follows:
Temperatures below 0
 °C:
R
T
 = R
0
[1 + A 
× T + B × T
2
 + C 
× T
3
 
× (T – 100 °C)]
Temperatures above 0 
°C:
R
T
 = R
0
[1 + A 
× T + B × T
2
]
T = temperature in 
°C
R
T
 = RTD resistance at temperature T
R
0
 = RTD nominal resistance at 0 
°C
A, B, C are coefficients given in Table 2.
Table 2 lists the coefficients used in this equation for each of the 
TCR values that the [c]FP-RTD-122 supports. If you have a 
nonstandard RTD that does not match one of these linearization 
curves, measure the resistance with the [c]FP-RTD-122 and 
convert the resistance to temperature in the manner suggested by 
the RTD vendor.
Three-Wire Compensation of Lead 
Resistance Errors
The [c]FP-RTD-122 uses a three-wire compensation technique to 
compensate for the lead resistances. The SENSE lead measures the 
resistance of the return COM lead. If the EX+ lead has the same 
resistance as the COM lead, the [c]FP-RTD-122 corrects for the 
Table 2.  Callendar-Van Dusen Coefficients Used by the [c]FP-RTD-122
TCR 
m
Ω/Ω/°C
A
(
°C)
–1
B
(
°C)
–2
C
(
°C)
–4
3.750
a
3.81 
×
 10
–3
–6.02 
×
 10
–7
–6.0 
×
 10
–12
3.851
b
3.9083 
×
 10
–3
–5.775 
×
 10
–7
–4.183 
×
 10
–12
3.911
c
3.9692 
×
 10
–3
–5.8495 
×
 10
–7
–4.233 
×
 10
–12
3.916
d
3.9739 
×
 10
–3
–5.870 
×
 10
–7
–4.4 
×
 10
–12
3.920
e
3.9787 
×
 10
–3
–5.8686 
×
 10
–7
–4.167 
×
 10
–12
3.928
f
3.9888 
×
 10
–3
–5.915 
×
 10
–7
–3.85 
×
 10
–12