Konica Minolta 752 用户手册

下载
页码 87
4-3.How To Operate Maintenance Mode
DRYPRO MODEL 752 Operation Manual Ver.0.11 2002.6    25
Automatic calibration
The DRYPRO has the automatic density calibration function
that perform the density calibration automatically after a lapse
of the specified time or specified number of printing. (Instead of
performing automatic calibration, you can make [CALIBRA-
TION] appear on the screen to show you that the calibration
needs performing depending on the settings at installation.)
Or the automatic density calibration can be performed when the
film is loaded.
CAUTION
• Even after a lapse of the specified period or
specified number of printing, the automatic
calibration is not performed if printing data is
not queued. The automatic calibration is per-
formed when the printing data is queued after
a lapse of the specified period or specified
number of printing.
• The automatic calibration is performed if the
printing data is not queued even when the film
is loaded. The automatic calibration is per-
formed when the printing data is queued on
the next time.