Chapter 1
Installing and Configuring the SCXI-1127/1128
© National Instruments Corporation
1-21
SH9MD-9MD cable
8-position HVAB plug
HV8-BAN4 cable
External DMM (if needed)
BAN2-WIRE kit (two kits needed for 4-wire), if not using the 
HV8-BAN4 cable
Cable or wires to connect the handshaking lines to the 
SCXI-1127/1128
1/8 in. flathead screwdriver
Consult the documentation for your SCXI chassis and other SCXI modules 
for additional instructions and warnings. You should have installed your 
SCXI-1127/1128 modules according to the instructions given in 
 section. Any other 
SCXI modules should already be installed according to their installation 
instructions. To install the SCXI-1127/1128 module with an SCXI chassis, 
follow these steps:
1.
Install the SCXI-1127/1128s in the chassis and install the HVAB as 
described in one of the Connecting the SCXI-1127/1128 Backplane 
sections.
2.
To connect an SCXI-1127/1128 to an external DMM device, you need 
two cables—the HV8-BAN4 or BAN2-WIRE cable, and a cable to 
connect the handshaking lines from the SCXI-1127/1128 to the DMM. 
Attach your HV8-BAN4 cable to the HVAB connector on the 
HVAB-backplane adapter or connect your BAN2-WIRE cable to the 
OUT0 terminals on the SCXI-1331 terminal block. When making 
four-wire measurements you must also connect a BAN2-WIRE cable 
to the OUT2 terminals.
3.
Attach the other end of the HV8-BAN4 or BAN2-WIRE cable to the 
inputs of your external DMM device.
4.
Connect the handshaking lines from the front of the SCXI-1331 
terminal block to the appropriate lines on the external DMM. See 
Chapter 2
, for information on how to use the 
SCXI-1127/1128 handshaking lines.
5.
Check the installation.
6.
Turn on the SCXI chassis.
7.
Turn on the computer or reconnect it to your chassis.
The SCXI-1127/1128 module is now installed.