5. 
After processing of film is complete, record the 
date, time, and processor identification number on 
the film in the designated areas. 
Manual Data Recording 
Procedure 
Establishing Normal Speed Index, Contrast 
Index, and Base+Fog 
The normal speed index, contrast index, and base+fog 
values are established on a representative film, when 
the processor is considered to be operating in an 
optimum fashion. 
Run several film samples and determine the average 
values for speed index, contrast index, and base+fog, 
using a transmission densitometer. Step wedge areas are 
as uniform as is possible to produce. There are, 
however, some errors at the edges of the step area. 
Therefore, always measure density at the center of 
the step
. Use a 2mm aperture to give best repeatability. 
Record the following data on the process control 
record. 
• 
Developer Temperature - Temperature of 
developer solution in processor during processing 
• 
Normal Base+Fog - Step number one density, the 
least exposed step on the wedge. 
• 
Normal Speed Index - The density of the step 
closest to 1.0D + base+fog. The step number 
should remain the same for a given processor and 
film type. 
• 
Normal Contrast Index - Select the step closest 
to but not larger than 2.20D. Subtract from this 
step the step closest to but not lower than 0.45D.