Rite Life 394 用户手册

下载
页码 28
NOTE: Monitor subsequent films on the same steps 
selected for normal contrast index. 
• 
Date - Month - Day – Year 
• 
Processor Number - Processor identification 
• 
Emulsion Number - Film batch identification 
• 
Developer Type - Developer vendor 
identification 
• 
Fixer Type - Fixer vendor identification 
• 
Film Type - Film vendor identification 
• 
Exposure Color - Exposure light (blue or green) 
• 
Developer Replenisher Rate - The rate of 
developer replenishment 
• 
Fixer Replenisher Rate - The rate of fixer 
replenishment 
• 
Processing Time - Film process time, input-to-
output 
A box of film should be set aside from regular stock for 
exclusive sensitometer use. New film stock will require 
re-establishment of normal values because small density 
changes are possible between film batches. 
Daily Plotting of Data on Process Control 
Record 
Plot the Speed Index, Contrast Index, Base+Fog, and 
Developer Temperature
 results on the process control 
record each time a control film is developed. Record 