Serial Communications Port (AUX)
♦ NI 5102 (PXI)
The serial communication port, AUX, provides +5V and GND for
applications that may require up to 100 mA of current operation and PF12
for triggering.
PFI2 has the same functionality as described above, but it is overloaded on
TRIG0 (SCANCLK) on the mini-DIN connector and is accessible only
through the optional 9-pin mini-DIN to BNC female cable adapter.
Analog Input
The two analog input channels are referenced to common ground in bipolar
mode. These settings are fixed; therefore, neither the reference nor the
polarity of input channels can be changed. You cannot use CH0 or CH1 to
make differential measurements or measure floating signals unless you
subtract the digital waveforms in software. For accurate measurements,
make sure the signal being measured is referenced to the same ground as
your NI 5102 by attaching the probe’s ground clip to the signal ground.
Table 4-2 shows the input ranges available on CH0 and CH1.
Note
The X10, X100, and X1000 designations are used to indicate a signal attenuation
rather than amplification. For example, with a X100 probe and a gain of 1, if you measure
a 400 V signal, the NI 5102 will receive 4 V (400 V/100 = 4 V) at its input connector.
The TRIG channel has a fixed input range of ±5 V. All NI 5102 digitizers
power up with a default gain of 1, thereby allowing the largest input range
available. TRIG channel range values are the same as the gain of 1 values
in Table 4-2.
Table 4-2. CH0 and CH1 Input Ranges
Gain
Input range
X1 Probe
X10 Probe
X100 Probe
X1000 Probe
1
±5 V
(default setting)
±50 V
±500 V
±5000 V
5
±1 V
±10 V
±100 V
±1000 V
20
±0.25 V
±2.5 V
±25 V
±250 V
100
±50 mV
±0.5 V
±5 V
±50 V