National Instruments cFP-RTD-122 用户手册

下载
页码 15
FP-RTD-122 and cFP-RTD-122
6
ni.com
Figure 5.  Four-Wire RTD Connections on One Channel
Taking Measurements from Two-Wire RTDs
Connect either wire of a two-wire RTD to the EX+ terminal and the 
other wire to the COM terminal, and connect a short jumper wire 
between the COM and SENSE terminals.
Figure 6.  Two-Wire RTD Connections on One Channel
Measuring Resistance Directly
You can use the [c]FP-RTD-122 to measure resistance in ohms. 
In this way, you can take measurements from RTDs of types that 
the [c]FP-RTD-122 does not directly support (such as 120
Ω nickel 
RTDs) and from resistive devices other than RTDs. You can choose 
one of two resistance ranges: 0–400 and 0–4,000
Ω. Resistance 
values outside the range you select, including open circuits, result 
in an 
Out of range
 error for the affected channels. The 
[c]FP-RTD-122 ignores any configuration of RTD type for 
channels with resistance ranges selected.
4-wire RTD
[c]FP-RTD-122
Leave this wire 
unconnected
EX+
SENSE
COM
2-wire RTD
EX+
SENSE
COM
[c]FP-RTD-122
Short Jumper Wire