FP-RTD-122 and cFP-RTD-122
Converting Resistance Measurements 
to Temperature Measurements 
The [c]FP-RTD-122 has built-in linearization algorithms for 
platinum RTDs of either 100 or 1,000 
Ω nominal resistance, 
and for six TCR (or alpha,
 
α) values. The TCR is the average 
temperature coefficient of resistance of an RTD from 0
 
to 100 
°
C. 
This document specifies TCR in units of m
Ω/Ω/°C.
The [c]FP-RTD-122 linearizes resistance values and returns 
readings in units of temperature. The available ranges are 
73 to 1,123 K, –200 to 850
°C, and –328 to 1,562 °F. You can 
configure each channel independently, so you can connect different 
types of RTDs to each channel.
Note
You must configure each channel of the 
[c]FP-RTD-122 for the RTD type connected to it. The 
module does not automatically recognize RTD types.
RTD Types
RTD types are specified by material composition, nominal 
resistance at 0
°C, and TCR. The [c]FP-RTD-122 can directly 
measure the temperature of platinum RTDs of either 100 or 
1,000
Ω nominal resistance. These RTDs are commonly referred to 
as PT100 or PT1000 RTDs, respectively.
Different types of platinum RTDs have different TCRs. The 
[c]FP-RTD-122 supports the following TCRs: 3.750, 3.851, 3.911, 
3.916, 3.920, and 3.928 m
Ω/Ω/°C. The most common TCR for 
RTDs is 3.851 m
Ω/Ω/°C. It is defined in international standards 
such as IEC-751, DIN 43760, BS 1904, and ASTM E1137. 
The TCR of 3.928 m
Ω/Ω/°C is used in the reference function for 
platinum thermometers in the International Temperature Scale 
of 1990 (ITS-90) for high-accuracy metrology applications. 
Unfortunately, not all TCR values are as well defined by standards 
organizations, and the behavior of RTDs with the same TCR value 
may vary from vendor to vendor. The variations are usually small, 
and the built-in linearization algorithms of the [c]FP-RTD-122 are 
appropriate for nearly all applications.