6.5.5  Manual Defect Correction 
 
 
Execute Manual Defect Detection when the defect pixels are visible to the naked eye 
even the automatic defect correction is processed. Manual Defect Detection is divided 
into the pixel type and line type. 
 
Manual Defect Correction in Pixel Type 
1  Make sure to carry out the automatic defect correction from 1 to 14 steps first. 
2  Click Show and Apply buttons on the left top of Manual Defect Detection
3  Press and hold the right mouse button and move it to left/right/top/ bottom in the FLAT image to 
adjust the brightness until you can verify the image with the naked eyes. 
4  Click Add Pixel Button. 
5  Use the Zoom or Pan function properly to find and click the location of defect pixel. Red point will be 
displayed and coordinate of the pixel will be added on the Detect Map List on the left upper. 
6  Click Save and Close button to save the Defect Map data with a new name. 
 
 
 
If a selected pixel has a difference in the brightness or gray level value (more than 15%) 
compared to the adjacent pixels, the pixel may be regarded as a defect pixel. 
 
A window located on the left top will magnify and display the pixel where the mouse 
pointer is located. The coordinate and Gray level value of the pixel will be displayed 
under the window.