Vieworks Co. Ltd. FXRD-1012NAW 用户手册

下载
页码 151
 
 
Rev.1.0
 
Page 112 of 151 
VW40-152-009
 
VIVIX-S 1012N User Manual 
 
Manual Defect Correction in Line Type 
1  Make sure to carry out the automatic defect correction from 1 to 12 steps first. 
2  Click Show and Apply buttons on the left top of Manual Defect Detection
3  Press and hold the right mouse button and move it to left/right/top/bottom in the FLAT image to 
adjust the brightness until you can verify the image with the naked eyes. 
4  Find and move to the start point of defect lines (line type) on the image. 
5  Click Add V-Line or Add H-Line button. 
6  Press and hold the left mouse button on the first defect pixel, move the mouse to the opposite 
direction and release the mouse button on the last pixel of Line Defect pixel. 
7  A red line will be displayed on the image and check if pixels on the line are added to the Defect Pixel 
list. 
8  Click Save and Close button to save the newly generated Defect Map data with a new name. 
 
 
 
 
 
If a selected pixel has a difference in the brightness or gray level value (more than 15%) 
compared to the adjacent pixels, the pixel may be regarded as a defect pixel. 
 
A window located on the left top will magnify and display the pixel where the mouse 
pointer is located. The coordinate and Gray level value of the pixel will be displayed 
under the window. 
 
 
 
If there are pixels existed without defect correction even though you conduct the manual 
defect correction, get technical support from the person in charge of Vieworks or a 
certified engineer by Vieworks.