Vieworks Co. Ltd. FXRD-1012NAW 用户手册

下载
页码 151
 
 
Rev.1.0
 
Page 113 of 151 
VW40-152-009
 
VIVIX-S 1012N User Manual 
 
Functions of Manual Defect Detection 
Function 
Description 
Show 
Decides whether to indicate the selected defect on the image 
Apply 
Decides whether to apply Defect in the list to the calibrated image. 
All 
Applies to all coordinates of defects added to the list 
Selected 
Applies to relevant coordinates of the selected defects from the list only. 
Auto Defect Finder 
Performs the auto defect correction. 
Delete 
Deletes the selected items from the list. 
Add Pixel 
Adds one defect pixel to Defect Map List. 
Add V-Line 
Adds vertical line type of defect pixels to Defect Map List 
Add H-Line 
Adds horizontal line type of defect pixels to Defect Map List. 
Delete Pixel 
Deletes the defect pixel of a point on list of Defect Map. 
Delete V-Line 
Deletes the defect pixel of vertical line on list of Defect Map. 
Delete H-Line 
Deletes the defect pixel of horizental line on list of Defect Map. 
ROI 
Marks the area of interest. 
Zoom 
Zoom in or out the image. 
W/L 
Adjusts the window level of the image to change the brightness. 
Pan 
Moves the image to a desired location. 
1:1 
Displays the image as 1:1 ratio. 
Fit 
Displays the image as fitting it on the screen. 
Save and Close 
Saves the changed defect pixel information as Defect map data. 
Close 
Closes the window without saving the changed defect pixel information