Vieworks Co. Ltd. FXRD-1012NAW 用户手册
Rev.1.0
Page 113 of 151
VW40-152-009
VIVIX-S 1012N User Manual
Functions of Manual Defect Detection
Function
Description
Show
Decides whether to indicate the selected defect on the image
Apply
Decides whether to apply Defect in the list to the calibrated image.
All
Applies to all coordinates of defects added to the list
Selected
Applies to relevant coordinates of the selected defects from the list only.
Auto Defect Finder
Performs the auto defect correction.
Delete
Deletes the selected items from the list.
Add Pixel
Adds one defect pixel to Defect Map List.
Add V-Line
Adds vertical line type of defect pixels to Defect Map List
Add H-Line
Adds horizontal line type of defect pixels to Defect Map List.
Delete Pixel
Deletes the defect pixel of a point on list of Defect Map.
Delete V-Line
Deletes the defect pixel of vertical line on list of Defect Map.
Delete H-Line
Deletes the defect pixel of horizental line on list of Defect Map.
ROI
Marks the area of interest.
Zoom
Zoom in or out the image.
W/L
Adjusts the window level of the image to change the brightness.
Pan
Moves the image to a desired location.
1:1
Displays the image as 1:1 ratio.
Fit
Displays the image as fitting it on the screen.
Save and Close
Saves the changed defect pixel information as Defect map data.
Close