7.3
Fault-finding
Symptom
Cause
Action
Poor probe repeatability 
and/or accuracy.
Debris on part or stylus.
Clean part and stylus.
Poor tool change repeatability.
Redatum probe after each tool 
change.
Loose probe mounting on shank 
or loose stylus.
Check and tighten as appropriate.
Calibration out of date and/or 
incorrect offsets.
Review probing software.
Calibration and probing speeds 
not the same.
Review probing software.
Calibration feature has moved.
Correct position.
Measurement occurs as stylus 
leaves surface.
Review probing software.
Measurement occurs within 
the machine’s acceleration and 
deceleration zone.
Review probing software and 
probe filter settings.
Probing speed too high or too 
slow.
Perform simple repeatability trials 
at various speeds.
Temperature variation causes 
machine and workpiece 
movement.
Minimise temperature changes.
Machine tool faulty.
Perform health checks on 
machine tool.
RMP60 status LEDs do 
not correspond to RMI 
or RMI-Q status LEDs.
Radio link failure – RMP60 out of 
RMI or RMI-Q range.
Check position of RMI or RMI-Q, 
see operating envelope.
RMP60 has been enclosed/
shielded by metal.
Change batteries.
RMP60 and RMI or RMI-Q are 
not partnered.
Partner RMP60 and RMI or 
RMI-Q.