Olympus Winter & Ibe GmbH ESG200 用户手册
80
Troubleshooting
Problem
Cause
Remedy
Foot switch pedal or
hand switch of the HF
instrument is pressed
and the output tone
sounds but no output
power is delivered.
The foot switch is assigned to another
output socket.
Check the correct assignment of the
foot switch. Refer to the section “Foot
switch assignment and Autostart” on
page 41.
Improper connection of the HF
instrument plug with the output socket
of the electrosurgical generator.
Check the plug of the HF instrument
for correct connection.
Malfunction of the electrosurgical
generator.
Do not use this electrosurgical
generator.
Contact Olympus.
No output power
is delivered when
an RFITT mode is
selected. The end-
of-procedure signal
sounds immediately
after activation.
The electrode has no contact with the
tissue.
Check that the electrode has contact
with the tissue.
Improper connection of the HF
instrument's plug with the output
socket of the electrosurgical
generator.
Check the plug of the HF instrument
for correct connection.
The HF instrument or its connection
cable is damaged.
Replace the HF instrument or its
connection cable.
Malfunction of the electrosurgical
generator.
Do not use this electrosurgical
generator.
Contact Olympus.
It is impossible to stop
the output power of
the electrosurgical
generator.
Autostart is assigned to the currently
used output socket and both
electrodes touch the tissue.
Remove the electrodes of the HF
instrument from the tissue.
Malfunction of the foot switch or hand
switch.
Immediately switch off the
electrosurgical generator by pressing
the power switch. Replace the foot
switch or the HF instrument with hand
switch.
Malfunction of the electrosurgical
generator.
Do not use this electrosurgical
generator.
Contact Olympus.
Its impossible to switch
off the electrosurgical
generator.
Malfunction of the electrosurgical
generator.
Disconnect the cord set plug from
the socket on the rear panel of the
electrosurgical generator and/or from
the grounded mains electricity socket.
Contact Olympus.
Do not use this electrosurgical
generator.
Table: Troubleshooting – General problems
12.2 Error messages
The electrosurgical generator is equipped with an intelligent alarm system. The alarm
system determines alarm conditions on the base of multiple variables.