Programming Guide
 
11.6 Device Control 
11.6.1 Output for the Target Temperature 
The signal output can be set to 4 – 20 mA, 0 – 20 mA or mV. If current 
output is activated, the output can provide a predefined current: 
XO=4 
output mode to 4 – 20 mA 
O=13.57 
output of a constant current at 13.57 mA 
O=60 
switches back to the temperature controlled output 
11.6.2 Analog Output, Scaling 
According to the temperature range of the model, it is possible to set 
a  maximum  voltage/current  value  according  to  a  temperature  value 
(e.g., the maximum current 20 mA shall represent 200°C / 392°F). The 
same setting is possible for the minimum value. 
H=500 
the maximum current/voltage value is set to 500°C 
L=0 
the minimum current/voltage value is set to 0°C 
Remark:  You  cannot  set  this  value  for  thermocouple  output.  The 
minimum span between the maximum / minimum settings is 20 K. 
11.6.3 Alarm Output 
The second output channel can be set in different modes, see section 
5.3.2 
 on page 20. 
•  Internal sensing head temperature 
•  Alarm output 
K=7 
internal sensing head temperature 
K=4 
alarm  output  for  object  temperature,  0  V  in  case  of  no 
alarm 
XS=125.3  threshold setting to 125.3°C (if U=C is set) 
11.6.4 Factory default values 
It is possible to reset the unit to the default values. 