Programming Guide 
XF 
factory default values will be set 
11.6.5 Lock Mode 
The  access  to  the  unit  is  possible  via  serial  interface  (software)  and 
via  the  direct  user  input  (mode  buttons,  LCD  display).  It  is  possible 
to lock the buttons. This allows access the unit only via software. 
J=L 
direct user input via mode buttons denied 
Alternatively  the  unit  can  be  unlocked  by  pressing  the  <Mode/Up> 
buttons simultaneously for three seconds. 
11.6.6 Mode Setting for the Digital Input FTC3 
The  digital  input  FTC3  (see  section  5.4.4 
 
on page 28) can be used as follows: 
XN=T 
FTC3 as trigger 
XN=H 
FTC3 with hold function 
11.6.7 Changing the Sensing Head Calibration Data 
If  it  is  necessary  to  exchange  the  sensing  head,  you  must  set  the 
calibration data for the new sensing head: 
XZ=0123 4567 FFFF FFFF 
according to the head calibration data 
For MID/MIC‐models the last eight  numbers are not used and must 
be set to „F“ like shown in the example above. For MIH‐models and 
specials (like MTB or G5) all sixteen numbers are used. 
11.6.8 Ambient Background Temperature Compensation 
In  case  of  compensating  the  ambient  background  temperature,  the 
following modes are available: 
AC=0 
no compensation 
AC=1 
compensation  with  a  constant  temperature  value  set 
with command A. 
MID 
73