5    Measuring 
    Setup of the Serial Interface 
   
All signs have to be in capital letters and shall be sent and received as an ASCII code. 
   Setup: 
    
Baud 
Rate 
9600 
 
    
Data 
bit 
 
    
Stop 
bit 
 
    
Parity 
None 
 
    
Handshake  XON/OFF 
 

   
Pin Assignment of PLC Interface XS 5 (97123) and PLC Interface XS 9 (97103) 
see Section 8.2 
   Selection Command 
   
Before you start the desired dispensing cycle you have to select the appropriate 
reference. 
   
The command you have to send is “X” and the number of the reference. 
   
For example, you send 
   
 
X0 
= Reference 0 is active 
 
   
 
X1 
= Reference 1 is active 
 
   
 
X2 
= Reference 2 is active 
 
   
 
X3 
= Reference 3 is active 
 
   
 
: : 
 
   
 
: : 
 
   
 
X9 
= Reference 9 is active 
 
   
The answer you get is “ok”. This means you can start your desired dispensing cycle. 
   
If you get an Error message, send the command again. In case of further Error 
messages see Chapter 7. 

   
Multiple choice of reference dispensing is not possible. 
   
If you need support, please ask Loctite for quotation on installation.