Operator's Manual  ZM-920PA/930PA
Insufficient Light Alarm
When sufficient light cannot be obtained from the photo emitter of the probe, the
transmitter generates the following sound and indication.
(1) “peep” sound (every 1 second)
(2) pulse level bar graph is maximum
(all bars are lit.)
(3) SpO
2
 data is displayed as “
(4)  %SpO
2
 is blinking.
In this case, change the attachment site to the appropriate site.  Refer to the
operator’s manual of the SpO
2
 probe.
Probe Malfunction Alarm
CAUTION
When any of the following occurs, the probe may be broken.
Replace it with a new one and check the probe.
••••• The transmitter generates “pip” sound every 0.25 seconds.
••••• SpO
2
 data is 85% and blinking.
When Measurement Condition is Unstable
SpO
2
 data blinks every 1 second when SpO
2
 signal stability decreases and the
transmitter cannot detect correct pulse waveform because of patient body
movement, poor attachment condition or poor circulation condition at the probe
attachment site.
CAUTION
SpO
2
 data blinking every second indicates an unstable pulse
waveform and displayed SpO
value may be inaccurate.  The
displayed data may not reflect sudden SpO
changes.