Nihon Kohden Corporation ZM-930PA Benutzerhandbuch

Seite von 70
Operator's Manual  ZM-920PA/930PA
43
Sound Display 
Cause  Countermeasure 
Single 
“peep” sound   
for 4 s 
--- 
The CALL key is pressed. The sound lasts 
while the key is pressed. 
--- (1) 
The 
battery 
voltage 
decreases and battery 
charge is almost zero.
Replace the 
batteries with new 
ones. 
Continuous 
“peep” sound   
All lights are 
off 
Battery is completely 
discharged. 
Replace the 
batteries with new 
ones. To stop the 
sound, turn off the 
power. 
Electrode lead is 
disconnected from the 
electrode. 
Firmly connect the 
electrode lead to 
the electrode. 
Electrode lead is 
disconnected from the 
transmitter. 
Firmly connect the 
electrode lead to 
the transmitter. 
Electrode lead 
discontinuity 
Replace the 
electrode lead with 
a new one. 
Electrode is not 
firmly attached to the 
skin. 
--- (2) 
Polarization voltage 
is abnormally high. 
Replace the 
electrode with a 
new one. 
Intermittent 
“pip” sound 
every 0.5 s 
--- SpO
2
 measurement 
site is under 
fluorescent light, 
surgical light, 
sunlight, etc. 
Cover the 
measurement site 
with a blanket or 
cloth. 
(1)
Displayed as LED on ZM-920PA
RA LA LL Va Vb
Alarm List
(2)
(5) %SpO
2
 indication
(4) pulse level bar graph
(3) SpO
2
 data