Operator's Manual  ZM-940PA
If the problem still remains after checking the following, contact your Nihon Kohden distributor.
Transmitter
Problem Cause 
Countermeasure 
Batteries are not installed 
correctly. The battery 
polarity is wrong. 
Install the batteries correctly. 
Nothing is 
displayed on the 
LCD after 
turning the 
power on. 
Batteries are completely 
discharged. 
Replace the batteries with new ones. 
LCD is difficult 
to see (too dark 
or too light). 
LCD brightness is not 
appropriate. 
Change the LCD brightness on the 
SYSTEM SETUP screen. Refer to the 
“Changing System Setup Settings” 
section. 
The channel of the 
transmitter and monitor 
does not match. 
Set the correct channel on the monitor. 
Nothing is 
displayed on the 
monitor after 
turning the 
transmitter 
power on. 
The software version of 
the multiple patient 
receiver is old. 
Upgrade the multiple patient receiver 
software to receive signal from the 
transmitter. The software version must be 
01-09 or later. 
Another transmitter of the 
same channel is used 
nearby. 
Turn the transmitter power off. If the 
monitor still receives a signal, there is a 
high probability that another transmitter 
of the same channel is used nearby. 
Follow the instruction of your channel 
administrator and use another transmitter 
of a different channel. 
Signals are mixing. 
Follow the instructions of your channel 
administrator and use another transmitter 
of a different channel. 
Signal receiving 
condition is poor. 
The transmitter is 
damaged. 
Contact your Nihon Kohden 
representative. 
 
Troubleshooting