Nihon Kohden Corporation ZM-940PA Benutzerhandbuch

Seite von 111
72
Operator's Manual  ZM-940PA
NIBP
Problem Cause 
Countermeasure 
The cuff hose is not connected 
to the NIBP socket properly. 
Connect the cuff hose to the socket 
properly. 
Cuff inflation 
pressure is less than 
10 mmHg. 
The cuff is not wrapped 
around the arm or is wrapped 
too loosely. 
Wrap the cuff around the upper 
arm. 
The cuff hose is not connected 
to the NIBP socket. 
Connect the cuff hose to the socket 
firmly. 
The cuff does not 
inflate when the 
NIBP START/STOP 
key is pressed. 
The cuff hose or extension 
hose may be folded or 
squeezed when the cuff 
pressure display on the screen 
increases quickly but the 
actual cuff does not inflate. 
Check the cuff hose and air hose. 
The cuff size is not correct. 
Select the cuff which fits the 
patient’s limb circumference. 
The cuff is not wrapped 
around the arm correctly. 
Wrap the cuff around the upper 
arm, not too tightly or too loosely. 
NIBP data is not correct 
because of body movement. 
Prevent the patient from moving 
during measurement. 
Check that nothing is touching the 
cuff during measurement. 
Abnormal 
measurement results 
are displayed. 
Vibration on the cuff. 
Change the measuring site. 
The cuff is suddenly 
deflated during 
inflation. 
The NIBP START/STOP key 
is pressed during inflation. 
 
Auto mode 
measurement does 
not start even when 
the time interval has 
passed. 
The NIBP INTERVAL key is 
pressed and the measurement 
mode is changed. 
Check the measurement mode and 
interval. 
The cuff suddenly 
inflates. 
The measurement mode is set 
to auto mode. 
Check the time interval. If 
necessary, stop measurement. 
Cannot connect cuff 
to the air hose. 
Unspecified cuff is used. 
Use a cuff specified by Nihon 
Kohden.