Nihon Kohden Corporation ZM-940PA Benutzerhandbuch

Seite von 111
Operator's Manual  ZM-940PA
71
ECG/Respiration
Problem Cause 
Countermeasure 
The heart rate is 
unstable. 
Pacing detection setting on the 
monitor is not correct. 
Turn off the pacing detection setting 
on the monitor. 
When monitoring a pacemaker 
patient, turn on pacing detection. 
Electrode lead is disconnected 
from the electrode. 
Firmly connect the electrode lead to 
the electrode. 
Electrode lead discontinuity 
Replace the electrode lead with a new 
one. 
Electrode is not firmly 
attached to the skin. 
Replace the electrode with a new one. 
The “CHECK 
ELECTRODE” 
message appears 
on the receiving 
monitor. 
Polarization voltage is 
abnormally high. 
Use Nihon Kohden specified 
electrodes. 
The gel on the electrode is 
dried out. 
The gel on the electrode is 
coming off. 
Replace the electrode with a new one. 
Electric blanket is used. 
Cover the blanket with a shield cover. 
ECG baseline is 
thick. 
(Hum is 
overlapping) 
Hum filter is set to OFF on the 
monitor 
Set the filter to ON. 
The gel on the electrode is 
dried out. 
Respiration 
waveform 
measurement is 
unstable. 
The gel on the electrode is 
coming off. 
Replace the electrode with a new one. 
 
SpO
2
Problem Cause 
Countermeasure 
The probe size is not 
appropriate for the patient. 
Use the appropriate probe for the patient. 
Probe attachment condition 
is poor. Probe is partly 
detached from the skin. 
External light gets in. 
Firmly attach the probe according to the 
procedure in the probe operator’s manual. 
Measurement site is dirty. 
Patient is wearing nail 
polish. 
Remove dirt and nail polish. 
When the probe and cuff are attached to 
the same limb, set “INHIBIT SpO
2
 
DURING NIBP” setting on the 
PARAMETER SETUP screen to ON.   
SpO
2
 data is 
unstable and 
not reliable. 
Probe is attached to the 
same limb that is used for 
NIBP measurement. 
Attach the probe to the opposite limb. 
Avoid a site where blood circulation 
condition changes greatly.