Nihon Kohden Corporation ZM-940PA Benutzerhandbuch
Operator's Manual ZM-940PA
73
Problem Cause
Countermeasure
Vibration on the cuff.
Check that nothing is touching the
cuff during measurement.
cuff during measurement.
The cuff hose or extension
hose is bent or squeezed.
hose is bent or squeezed.
Remove the cause.
Cannot measure
NIBP.
NIBP.
The cuff has worn out.
Use a new cuff.
Increase the measuring interval.
Increase the measuring interval.
Blood congestion
occurs.
occurs.
Measuring over a long
period of time at short
intervals.
period of time at short
intervals.
Do not measure NIBP over a long
time.
time.
Thrombus occurs.
Measuring on a patient
with known bleeding
disorders or coagulation.
with known bleeding
disorders or coagulation.
Do not perform NIBP measurement
on such a patient.
on such a patient.
NIBP data on the
screen is --- or dark.
screen is --- or dark.
The time set for “OLD
NIBP DATA” on the
PARAMETER SETUP
screen elapsed from the last
measurement.
NIBP DATA” on the
PARAMETER SETUP
screen elapsed from the last
measurement.
When NIBP is measured again, the
data is displayed in normal
brightness.
data is displayed in normal
brightness.
Three loud pip
sounds indicting
NIBP measurement
cannot be started.
sounds indicting
NIBP measurement
cannot be started.
The cuff is not deflated
enough to start another
measurement.
enough to start another
measurement.
Wait 30 seconds and measure again.