Nihon Kohden Corporation ZM-940PA Benutzerhandbuch

Seite von 111
Operator's Manual  ZM-940PA
73
Problem Cause 
Countermeasure 
Vibration on the cuff. 
Check that nothing is touching the 
cuff during measurement. 
The cuff hose or extension 
hose is bent or squeezed. 
Remove the cause. 
Cannot measure 
NIBP. 
The cuff has worn out. 
Use a new cuff. 
Increase the measuring interval. 
Blood congestion 
occurs. 
Measuring over a long 
period of time at short 
intervals. 
Do not measure NIBP over a long 
time. 
Thrombus occurs. 
Measuring on a patient 
with known bleeding 
disorders or coagulation. 
Do not perform NIBP measurement 
on such a patient. 
NIBP data on the 
screen is --- or dark. 
The time set for “OLD 
NIBP DATA” on the 
PARAMETER SETUP 
screen elapsed from the last 
measurement. 
When NIBP is measured again, the 
data is displayed in normal 
brightness. 
Three loud pip 
sounds indicting 
NIBP measurement 
cannot be started. 
The cuff is not deflated 
enough to start another 
measurement. 
Wait 30 seconds and measure again.