Ortivus AB 410000409 User Manual
70
10 Appendix
Database full. No new
data will be saved for
this patient.
The limit of 4 hours
storage of patient
data has been
exceeded.
Monitoring of real time data is still possible, however
no new data will be stored and thus cannot be sent to
an HWS. A new care session could be started in order
to store and transmit new data.
MIDA relearn failed.
Poor electrode
contact.
Check ECG signal quality in Electrode test view.
Check electrode contact, check if old electrodes,
check that ECG cable is not damaged. Then try MIDA
Relearn again.
No communication
with hospital unit due
to program error.
Please contact
support for more
information.
Program error.
Contact authorised personnel.
Monitor mode only
due to program error.
Please contact
support for more
information.
Program error.
Contact authorised personnel.
Facsimile failed.
Fax request did not
reach hospital unit.
Check communication status and try again.