Ortivus AB 410000409 User Manual
72
10 Appendix
10.4.3 Batteries
Problem
Cause
Action
The 300c batteries
can not charge.
The 300c does not
receive external
power.
Make sure that the two indicators for external power
supply connected and battery charging are lit, see
“Units in the system” on page 5.
300c is too cold or hot
to charge.
Warm or cool 300c in order to charge 300c battery.
The operating time of
the 300c batteries is
low.
Batteries are getting
old.
Leave to authorised personnel for replacing batteries.
The 300m batteries
do not charge.
The 300m does not
receive external
power.
Make sure that the 300m receives external power.
Make sure that the 300m is correctly docked.
The operating time of
the 300m is short.
Batteries are getting
old.
Leave to authorised personnel for replacing batteries.