A monitor that is not connected to AC power will not communicate
with the transmitter – monitor data port is disabled when not
connected to AC power – so the AlarmView System will not transmit
in case of a monitor power failure. The only exception is the Nellcor
N-395 monitor, which will continue to communicate with the
transmitter while operating on battery power.
Outlets used for the system should be on the hospital emergency
circuit so that they are energized at all times.
with the transmitter – monitor data port is disabled when not
connected to AC power – so the AlarmView System will not transmit
in case of a monitor power failure. The only exception is the Nellcor
N-395 monitor, which will continue to communicate with the
transmitter while operating on battery power.
Outlets used for the system should be on the hospital emergency
circuit so that they are energized at all times.
Transmitter Placement
1. Place transmitter on top of or beside pulse oximeter it is
being used with.
To minimize possibility of the transmitter creating interference
with the monitor, place transmitter at the rear of monitor. For best
results with N-395 monitor, place transmitter to the right rear of
monitor.
{Insert transmitter placement}
Preferred N-395 Transmitter Placement
Turning Transmitter On
!
!
!
The antenna must be connected to the transmitter before the
transmitter is turned on. Turning the transmitter on without the
antenna connected may damage the transmitter. Do not disconnect
the antenna while the transmitter is turned on.
transmitter is turned on. Turning the transmitter on without the
antenna connected may damage the transmitter. Do not disconnect
the antenna while the transmitter is turned on.
1. Check oximetry monitor to ensure it is on and functioning
properly, and that alarm settings are appropriate.