Data Critical Corporation AVTS1010 User Manual

Page of 55
AlarmView™ System User’s Guide 
 
 
 
{insert part #} Page 20 
 
2. Press On/Off button on top of transmitter. 
 
Indicators will flash on and off while transmitter performs a 
Power-On Self Test (POST) to check its function. 
 
{Insert transmitter front view with Tx and Mode LED labels} 
 
Transmitter LED Indicators 
 
POST Light Sequence 
 
Stage 
Stage Indication 
Test 
Failure 
Indication 
Tx LED Green 
Transmitter 
program flash 
memory check 
Simultaneous 
flashing of Tx 
and Mode LEDs 
Mode LED 
Green 
External SRAM 
Simultaneous 
flashing of Tx 
and Mode LEDs 
Tx LED Amber 
Operation of 
IrDA UART 
Simultaneous 
flashing of Tx 
and Mode LEDs 
Mode LED 
Amber 
RF board synch 
lock 
Alternate 
flashing of 1 Tx 
and 1 Mode 
LED 
Mode LED 
Amber 
Tx ID flash read  Alternate 
flashing of 2 Tx 
and 2 Mode 
LEDs 
Mode LED 
Amber 
Setup flash read  Alternate 
flashing of 3 Tx 
and 3 Mode 
LEDs 
 
!
!
!
 
 
 
Stage 6 Failure is indication of hardware failure, even if transmitter 
seems to recover (only exception is when installer has programmed 
new software into transmitter). In case of a Stage 6 failure, all