Troubleshooting
Error code Message
Cause
Remedial action
E0104
E0105
Footswitch pedal pressed
1. Release the footswitch 
pedal to continue.
2. If the problem persists, 
contact Olympus.
The foot switch pedal is 
depressed while switching 
on the electrosurgical 
generator.
Release the foot switch 
pedal.
Malfunction of the foot 
switch.
Replace the foot switch.
Malfunction of the 
electrosurgical generator.
Do not use this 
electrosurgical generator.
Contact Olympus.
E0108
E0109
Handswitch pressed
1. Release the handswitch 
of the instrument to 
continue.
2. If the problem persists, 
replace the instrument.
3. If the problem persists, 
contact Olympus.
The hand switch of the 
HF instrument is pressed 
while switching on the 
electrosurgical generator.
Release the hand switch of 
the HF instrument.
Malfunction of the HF 
instrument.
Replace the HF instrument.
Malfunction of the 
electrosurgical generator.
Do not use this 
electrosurgical generator.
Contact Olympus.
E0115
Application time limit 
exceeded
1. Release the footswitch or 
handswitch and reactivate 
to continue.
2. If the problem persists, 
contact Olympus.
The maximum time limit for 
the application has been 
exceeded.
Release the foot switch or 
hand switch and reactivate 
by pressing the foot switch 
or hand switch again.
Malfunction of the 
electrosurgical generator.
Do not use this 
electrosurgical generator.
Contact Olympus.
E0136
Double pedal footswitch 
not assigned
1. Assign the double 
pedal foot switch to the 
designated output socket.
2. If the problem persists, 
contact Olympus.
The double pedal foot 
switch is not assigned to the 
corresponding output socket.
Assign the double pedal foot 
switch to the corresponding 
output socket. Refer to 
the section “Foot switch 
assignment and Autostart” 
on page 41.
Malfunction of the 
electrosurgical generator.
Do not use this 
electrosurgical generator.
Contact Olympus.
E0140
No mode selected
1. Select a mode.
2. If the problem persists, 
contact Olympus.
No mode has been selected. Select an appropriate mode. 
Refer to the section “Output 
settings” on page 59.
Malfunction of the 
electrosurgical generator.
Do not use this 
electrosurgical generator.
Contact Olympus.
E0141
Power set to zero (--)
1. Set a valid power level.
2. If the problem persists, 
contact Olympus.
The output power level for 
the chosen mode is set to 0.
Increase the output power 
level on the Set Screen. 
Refer to section “Selecting 
the appropriate output 
settings” on page 62.
Malfunction of the 
electrosurgical generator.
Do not use this 
electrosurgical generator.
Contact Olympus.