Measurement Specialties 9116 User Manual

Page of 143
Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 63 
 
www.PressureSystems.com 
CALCULATE AND SET OFFSETS  (Command ‘h’) 
 
Purpose: 
Instructs a module to calculate new offset coefficients with zero differential  
  pressure 
(or 
specified “generated” pressure) applied to the specified channels.   
 
 
These new coefficients update part of the module’s internal calibration coefficient  
 
 
database, used to convert any subsequent raw data into engineering units data.   
 
 
The  new offset values are also returned in the response. This command is  
 
 
sometimes called a Re-zero or zero-only calibration. 
 
 
 
Command 
hpppp [vv.vvvv]” 
 
h’  is the command letter 
pppp’ is the  position field 
vv.vvvv
’ is an [optional] applied pressure value preceded by a space 
character 
 
Response 
g.gggg .. [g.gggg]...” 
 
g.gggg’ are the actual offset data values returned, each preceded by a  
space. 
 
 
Description:  The  position field may have 0 or 4 characters.  If no position field is specified,  
 
 
  offset coefficients for all of a module’s input channels will be calculated and  
 
 
  returned.  If a  position field is specified, offset coefficients for only the channels  
 
 
  whose bits are set (=1) will be calculated and returned. If the optional pressure  
  
 
value 
vv.vvvv] is specified, the position field must be 4 characters, even when  
 
 
  all channels are to be specified.  If the optional pressure value [‘vv.vvv’] is not  
 
 
  provided, an applied pressure of 0.0 psi(a) will be assumed when calculating  
 
 
  coefficients. Offset values are returned in the response in order of highest  
  
 
specified 
channel to lowest specified channel, with data formatted per an implied  
 
 
  decimal format (f=0). 
 
  Before acquiring data with this command, any addressed Model 9116 module  
 will 
normally attempt to place the calibration valve in the CAL position, so that a  
  zero differential pressure can be applied to all channels via the module’s CAL  
  and CAL Ref input port.  Simply leaving these ports unattached will allow the  
  transducers to read the appropriate zero differential pressure if ambient air  
  pressure is stable.  After data are acquired, the calibration valve will be placed in  
  the RUN position.  To disable the automatic shifting of the calibration valve, refer  
 to 
the 
Set Operating Options (‘w’) command (index 0B).  The reader is also  
 referred 
to 
Section 4.2 of Chapter 4 for additional details concerning the  
  performance of a Re-zero Calibration. 
 
Note
 
The calculated offset values from the latest ‘h’ command will be lost 
when the module is powered off.  To save these offset terms to each 
transducer’s non-volatile memory refer to the Set Operating Options
 
(‘w’) command (index 08).