Command ‘c’ — Sub-command Index 06: Select Protocol For Stream Delivery
 
 
 
 
Command 
c 06 st  pro  [remport  [ipaddr]]” 
 
‘c’  is the command letter. 
‘ 06’ is the sub-command index (ii) for Select Protocol
‘ st’ is the stream id digit (0=the ONLY acceptable entry). 
‘ pro’ is the protocol id digit (1=UDP/IP, 0=TCP/IP) 
‘ remport’ is an optional remote port number to which each UDP stream is 
directed in the host (port 9000 is the default if unspecified).  It is ignored if 
pro=0. 
 ipaddr’ is an optional host IP address to which each UDP stream is 
directed (default is the host IP address per current TCP connection that 
sent this command).  It is ignored if pro=0. 
 
NOTE: all parameters separated by a space. 
 
Response 
A” 
‘A’ is the acknowledge letter 
 
Description:  This command sets the protocol by which every configured autonomous stream  
 
 
  is delivered to the host.  It must be executed after streams are configured,  
 
 
  but before they are enabled. 
 
  By default, streams are delivered via the same TCP/IP protocol used to receive  
  commands from host  (i.e., via the existing TCP/IP connection used to send this  
  command).  However, for special circumstances, all autonomous streams may  
  be delivered to the host via the UDP/IP protocol instead. This command is  
  required only when UDP/IP  is to be used. It also can restore the default protocol  
  (to TCP/IP) once it has been changed.  The TCP/IP version of the command  
  ignores the optional (pro and ipaddr) parameters, which have meaning only to  
 the 
UDP/IP 
protocol. 
 
  Though the command has a stream parameter, it is currently limited to changing  
 the 
protocol 
of 
all defined streams at the same time (i.e., parameter st must be =  
 0, 
meaning 
all configured streams).   
 
  The 
optional 
remport parameter may be any value in the range 1024 to 65535.   
  However, 
remport = 7001 should be avoided, since NetScanner modules emit  
 
 
UDP Query responses to that port, and most host programs should have a UDP  
socket already bound to that port for receiving these special responses.  The  
choice of  remport parameter will affect the way host software must handle  
multiple modules sending streams.  If every module uses the same port, then a  
single host socket can be bound to that port to receive all responses from all  
modules.  The remote IP address, given to that socket, at time of receipt of the  
stream’s datagram, will identify the particular module who sent the datagram.   
Alternately, every module may be given a unique remport number, requiring that  
a host program bind a unique UDP socket to each of these unique remport  
numbers.  Then, when a particular socket receives a UDP datagram to its unique  
port, the module sending it is automatically identified.