Agilent Technologies N5700 User Manual

Page of 128
Measure Commands 
Series N5700 User’s Guide 
67 
CALibrate:STATe <Bool> [,<NRf>] 
CALibrate:STATe? 
This command enables/disables calibration mode. Calibration mode 
must be enabled for the power supply to accept any other calibration 
commands. The first parameter specifies the enabled or disabled 
state On (1) or Off (0). The second parameter is the password.  
A password is required if calibration mode is being enabled and the 
existing password is not 0.  If the password is not entered or is 
incorrect, an error is generated and the calibration mode remains 
disabled. The query returns only the state, not the password. 
The *RST value = Off. 
CALibrate:VOLTage[:LEVel] <NRf> 
This command initiates the calibration of the output voltage. The 
value that you enter selects the range that is being calibrated. 
 
 
Measure Commands 
Measure commands measure the output voltage or current. MEASure 
commands acquire new data before returning the reading. 
Measurement overflows return a reading of 9.91E+37.  
MEASure[:SCALar]:CURRent[:DC]? 
MEASure[:SCALar]:VOLTage[:DC]? 
These queries perform a measurement and return the DC output 
current in amperes or DC output voltage in volts.