Agilent Technologies N5700 User Manual

Page of 128
Source Commands 
Series N5700 User’s Guide 
69 
Source Commands 
Source commands program the voltage, current, triggered, and 
protection functions. 
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <NRf+> 
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?  
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] <NRf+> 
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]?  
These commands set the immediate and the triggered output current 
level. The values are programmed in amperes. The immediate level is 
the output current setting. The triggered level is a stored value that is 
transferred to the output when a trigger occurs. At *RST, the 
immediate and triggered current values = 0. 
[SOURce:]CURRent:PROTection:STATe <Bool> 
[SOURce:]CURRent:PROTection:STATe?  
This command enables or disables the over-current protection (OCP) 
function. The enabled state is On (1); the disabled state is Off (0). If 
the over-current protection function is enabled and the output goes 
into constant current operation, the output is disabled and OC is set 
in the Questionable Condition status register. The *RST value = Off. 
An over-current condition can be cleared with the Output Protection 
Clear command after the cause of the condition is removed.  
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <NRf+> 
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]?  
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <NRf+> 
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]?  
These commands set the immediate and the triggered output voltage 
level. The values are programmed in volts. The immediate level is the 
output voltage setting. The triggered level is a stored value that is 
transferred to the output when a trigger occurs. At *RST, the 
immediate and triggered voltage values = 0. 
The range of values that can be programmed for these commands is  
coupled with the voltage protection and the voltage limit low settings. 
The maximum value for the immediate and triggered voltage level is 
either the value in the following table, or the voltage protection 
setting divided by 1.05; whichever is lower. The minimum value is 
either the value in the table, or the low voltage setting divided by 
0.95; whichever is higher.  
Note that triggered values can be programmed outside these limits, 
but an error will be generated when the trigger occurs. 
 
Model 
(V 
rating)  6V 8V 12.5V 20V 30V 40V 60V 80V 100V 150V 300V 600V 
Min. 
voltage 
level 
 0 0 0 0 0 0 0 0 0 0 0 0 
Max. 
voltage 
level  6.3  8.4 13.125  21  31.5  41.9  62.85  83.8 104.76 157.1 314.2 628.5