Mindray DS USA Inc. WT0707-00000 Manual De Usuario

Descargar
Página de 484
Transfer Tab
Patient Setup Functions
5 - 24
0070-10-0690-01
Panorama™ Operating Instructions
• The transferred patient's parameter display waveform 2 changes to OFF when the 
destination device does not support the source device's waveform 2 parameter.
• If the source device is a bedside monitor, the units for CO
2
 and temperature will change 
to that of the destination device.
• If the destination device and the source device are different, the source devices sensor 
display setup will be reset to the system default settings of the destination device. 
• The care group is set to the system default at the destination device.
Demographics
• For all transfers to a bedside monitor, the bed ID and patient size change to those of the 
destination device. The remaining demographic data only changes when the destination 
device has experienced a Communications Lost event. When communication is 
re-established, the remaining demographic data is taken from the destination device.
• For all transfers to a Panorama Telepack, the bed ID is cleared and the patient size 
changes to Adult (the Telepack default). The remaining demographic data is preserved.
Alarms
• The transferred patient's Alarm Limits, Alarm Delay, and ST Alarm Delay change to those 
of the destination device.
• All alarms for the transferred patient become enabled, discontinuing all Alarm Suspended 
settings.
5.2.3
Central To Central Transfer Without Device
Central To Central Transfer Without Device is used to transfer patient data stored 
at one Panorama Central Station to a different Panorama Central Station. Historical patient 
data is preserved upon successful completion of the transfer, with the exception of historical 
failed page data. If Paging is enabled at the central station, a successful Central To Central 
Transfer Without Device will delete any rows in the Failed Pages tab for the transferred 
patient.
The Central To Central Without Device button is enabled under the following 
conditions:
• When a patient tile actively monitoring a patient is selected as the source
• When a patient tile actively monitoring a patient, but placed in Standby mode, is 
selected as the source
• When a patient tile actively monitoring a patient, but experiencing a Communications 
Lost event, is selected as the source
For information regarding changes that occur at the destination tile when the transfer is 
complete, refer to ‘‘Effects of the Transfer’’ on page 5-27.
5.2.3.1
Source Panorama
1. A prerequisite for the transfer is that certain patient demographic information must exist. 
In the patient tile to be transferred, ensure that the appropriate information has been 
provided in at least two of the three demographic fields (Last Name, First Name, and 
ID).