Monitor Setup
M310 Fall Monitor – Setup & User Guide
Monitor Setup
Pad Exit Alarm Tone Adjustments
Loosen the battery cover screw and remove the battery cover to reveal 
the settings switches and buttons.
Volume (Speaker icon)  
There are 4 Volume settings for the Pad Exit alarm and Trouble Alarm 
tones:
•  1 - Low volume
•  2 - Medium volume
•  3 - High volume
•  NC (Nurse Call) - If a Nurse Call jack is plugged in and this setting is 
selected, the local alarm sound is suppressed completely, and the 
alarm is sent to the Nurse Call station.
Tone Selection (Music Notes icon)  
There are 4 different Alert tones which can be selected:
•  1 - One sound (low) in a sequence of 3 beeps, pause, 2 beeps
•  2 - Two sounds (low high low high) in a sequence of 4 beeps, pause, 
2 beeps
•  3 - Two sounds (low high low) in a sequence of 3 beeps, pause, 2 
beeps
•  4 - Two sounds (high low high) in a sequence of 3 beeps, pause, 2 
beeps
Delay (Stopwatch icon)  
There are 3 delay settings before the alarm is sounds:
•  0 (0 - 0.75 seconds)
•  2 (2 - 2.75 seconds)
•  4 (4 - 4.75 seconds)
Custom Pad Exit Alarms
1. 
The Play and Record buttons and a microphone are found to 
the right of the Tone, Volume, and Delay switches.
2. 
Press and hold the Record button, beside the icon that looks like 
a microphone.