Hardware Features 
M310 Fall Monitor – Setup & User Guide 7
Yellow Status Indicator Icons
Yellow indicates equipment trouble, including low battery or 
supervision alarms when communication is lost. Yellow accompanied 
by an alarm indicates a problem that prevents the monitoring of a 
patient/resident on the pad.
Audible Alerts
The monitor provides three types of audible alerts: Pad Exit Alarm, 
Trouble Alarm, and Confirmation Tones.
Pad Exit Alarms
Pad Exit Alarms are sounded when a patient/resident ceases to be 
monitored by the Sensormat pad.
There are 4 alarm tone sequences associated with the red alarm LED 
flashes. The choice of alarm or trouble tone depends on the TONE 
switch setting. Confirmation tones and alert tones are not affected by 
these settings.
Trouble Alarms
Trouble alarms are indicated by a negative tone when system problems 
are detected. Trouble Alarm tones are sounded once every 5 seconds 
and associated with the status LED (yellow flashes):
•  One sound (low) in a sequence of 3 beeps
•  Two sounds (low high low high) in a sequence of 4 beeps
•  Two sounds (high low high) in a sequence of 3 beeps
Confirmation Tones
Confirmation Tones (positive or negative) are sounded to verify 
successful operations. These tones are much quieter than the Alarm 
tone and cannot be cancelled:
•  A positive confirmation tone indicates a successful association of 
monitor and pad, or on pad monitoring is arming.