EPA Calibration Protocol 
Teledyne API – Model T300/T300M CO Analyzer 
236 
 
10.2.1.  ZERO/SPAN CHECK PROCEDURES 
The Zero and Span calibration can be checked in a variety of different ways.  They 
include: 
  Manual Zero/Span Check - Zero and Span can be checked from the front panel 
touchscreen.  The procedure is in Section 9.3 of this manual. 
  Automatic Zero/Span Checks - After the appropriate setup, Z/S checks can be 
performed automatically every night.  See Section 9.3 of this manual for setup and 
operation procedures.   
 
If using the AutoCal feature to perform a calibration check, set the CALIBRATE 
parameter to NO
  Zero/Span checks via remote contact closure = Zero/Span checks can be initiated 
via remote contact closures on the rear panel.  See Section 9.3.3.3 of this manual. 
  Zero/Span via RS-232 port - Z/S checks can be controlled via the RS-232 port.  See 
Section 9.3.3.3 and Appendix A-6 of this manual for more details. 
 
10.2.2. PRECISION CHECK 
A periodic check is used to assess the data for precision.  A one-point precision check 
must be carried out at least once every 2 weeks on each analyzer at a CO concentration 
between 8.0 ppm and 10.0 ppm.  
The analyzer must be operated in its normal sampling mode, and the precision test gas 
must pass through all filters, scrubbers, conditioners, and other components used during 
normal ambient sampling.  
The standards from which precision check test concentrations are obtained must be 
traceable to NIST-SRM.  Those standards used for calibration or auditing may be used. 
To perform a precision check during the instrument set up, the sources of zero air and 
sample gas and procedures should conform to those described in Section 9.2 for 
analyzers with no valve options or with an IZS valve option installed and Section 9.3.1 
for analyzers with Z/S options installed with the following exception: 
  Connect the analyzer to a precision gas that has a CO concentration between 8.0 
ppm and 10.0 ppm.  If a precision check is made in conjunction with a zero/span 
check, it must be made prior to any zero or span adjustments. 
  Record this value.  
Information from the check procedure is used to assess the precision of the monitoring 
data; see CFR 40 CFR 58
5
 for procedures for calculating and reporting precision. 