3.3. CONNECTIONS AND SETUP 
This section presents the electrical (Section 3.3.1) and pneumatic (Section 3.3.2) 
connections for setup and preparing for instrument operation 
3.3.1. ELECTRICAL CONNECTIONS 
Note 
To maintain compliance with EMC standards, it is required that the cable 
length be no greater than 3 meters for all I/O connections, which include 
Analog In, Analog Out, Status Out, Control In, Ethernet/LAN, USB, RS-232, 
and RS-485.   
3.3.1.1. CONNECTING POWER 
Attach the power cord to the analyzer and plug it into a power outlet capable of carrying 
at least 10 A current at your AC voltage and that it is equipped with a functioning earth 
ground. 
 
 
WARNING 
E
LECTRICAL 
S
HOCK 
H
AZARD
 
High Voltages are present inside the analyzer’s case. 
Power connection must have functioning ground connection. 
Do not defeat the ground wire on power plug. 
Turn off analyzer power before disconnecting or  
connecting electrical subassemblies. 
Do not operate with cover off. 
 
 
CAUTION
 
G
ENERAL 
S
AFETY 
H
AZARD
 
The T300/T300M Analyzer can be configured for both  
100-130 V and 210-240 V at either 47 Hz or 63 Hz.   
To avoid damage to your analyzer, make sure that the AC power voltage matches 
the voltage indicated on the analyzer’s model/specs label (See Figure 3-4
) before 
plugging the T300/T300M into line power.
 
 