Teledyne API – Model T300/T300M CO Analyzer 
Getting Started 
47 
3.3.1.2. CONNECTING ANALOG INPUTS (OPTION)  
The Analog In connector is used for connecting external voltage signals from other 
instrumentation (such as meteorological instruments) and for logging these signals in the 
analyzer’s internal DAS.  The input voltage range for each analog input is 0-10 VDC. 
 
Figure 3-8:  Analog In Connector 
Pin assignments for the Analog In connector are presented in Table 3-4. 
Table 3-4: 
Analog Input Pin Assignments 
PIN DESCRIPTION 
DAS 
PARAMETER
1
 
Analog input # 1 
AIN 1 
Analog input # 2 
AIN 2 
Analog input # 3 
AIN 3 
Analog input # 4 
AIN 4 
Analog input # 5 
AIN 5 
Analog input # 6 
AIN 6 
Analog input # 7 
AIN 7 
Analog input # 8 
AIN 8 
GND 
Analog input Ground 
N/A 
1
 See Section 7 for details on setting up the DAS. 
3.3.1.3. CONNECTING ANALOG OUTPUTS 
The T300 is equipped with several analog output channels accessible through a 
connector on the back panel of the instrument.  The standard configuration for these 
outputs is mVDC.  An optional current loop output is available for each. 
When the instrument is in its default configuration, channels A1 and A2 output a signal 
that is proportional to the CO concentration of the sample gas.  Either can be used for 
connecting the analog output signal to a chart recorder or for interfacing with a 
datalogger. 
Output A3 is only used on the T300/T300M if the optional CO
2
 or O
sensor is installed.   
Channel A4 is special.  It can be set by the user (see Section 5.9.8.1) to output any one 
of the parameters accessible through the <TST TST> buttons of the units sample 
display. 
To access these signals attach a strip chart recorder and/or data-logger to the appropriate 
analog output connections on the rear panel of the analyzer.   