Getting Started 
Teledyne API – Model T300/T300M CO Analyzer 
48 
 
ANALOG OUT
        A1          A2            A3           A4
     +     -     +     -       +     -      +      -
 
Figure 3-9: 
Analog Output Connector 
Table 3-5: 
Analog Output Pin-Outs 
PIN 
ANALOG OUTPUT 
VOLTAGE SIGNAL 
CURRENT SIGNAL 
V Out 
I Out + 
A1 
Ground 
I Out - 
V Out 
I Out + 
A2 
Ground 
I Out - 
V Out 
I Out + 
A3  
(Only used if CO
2
  or 
O
Sensor is 
installed) 
Ground 
I Out - 
V Out 
I Out + 
A4 
Ground 
I Out - 
 
3.3.1.4. CURRENT LOOP ANALOG OUTPUTS (OPTION 41)  SETUP 
If your analyzer had this option installed at the factory, there are no further connections 
to be made. Otherwise, it can be installed as a retrofit for each of the analog outputs of 
the analyzer . This option converts the DC voltage analog output to a current signal with 
0-20 mA output current. The outputs can be scaled to any set of limits within that 0-20 
mA range. However, most current loop applications call for either 2-20 mA or 4-20 mA 
range.  
Figure 3-10 provides installation instructions and illustrates a sample combination of 
one current output and two voltage outputs configuration. Following Figure 3-10 are 
instructions for converting current loop analog outputs to standard 0-to-5 VDC outputs. 
Information on calibrating or adjusting these outputs can be found in Section 5.9.3.7 
 
 
CAUTION – AVOID INVALIDATING WARRANTY 
Servicing or handling of circuit components requires electrostatic 
discharge protection, i.e. ESD grounding straps, mats and containers.  
Failure to use ESD protection when working with electronic assemblies will 
void the instrument warranty. Refer to Section 14 for more information on 
preventing ESD damage. 