Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
Modify the Cal Kit Thru Definition 
With this method, it is only necessary to use a thru adapter. The calibration kit thru 
definition is modified to compensate for the adapter and then saved as a user kit. However, 
the electrical delay of the adapter must first be found. The adapter match will degrade the 
effective load match terms on both ports as well as degrade the transmission frequency 
response (tracking).
1. Refer to 
 while performing the steps in this procedure. Also refer to 
 for an explanation of A1, A2, and A3.
2. Perform a 1-port calibration at “Reference Port 1”. Refer to Step A of 
.
Figure 6-19 Determining the Electrical Delay Setup
3. Connect the A3 adapter to “Reference Port 1” as shown in Step B of 
Connect a short to the open end of the A3 adapter.
4. Measure the delay of the adapter by pressing 
 
.
5. Divide the resulting delay measurement by 2 to determine the delay of the thru and the 
short in one direction.
6. Determine the offset delay of the calibration short by examining the define standard 
menu (see 
7. Subtract the offset delay of the short (determined in step 6) from the delay of the thru 
and the short in one direction (determined in step 5). The result is electrical delay of the 
thru. This value is used in the next step.
8. Modify the calibration kit thru definition by entering in the electrical delay of adapter 
A3. Save this as a user kit.
For example, if A3 has 100 ps of delay, press:
 
 
 
 
 
 
 
 
 
 
 
 
Format
 DELAY 
Cal
 CAL KIT   MODIFY   DEFINE STANDARD 
4
x1
 MODIFY STD DEFINITION  SPECIFY OFFSET 
 OFFSET  DELAY 
0.100
G/n
 STD DONE (DEFINED)   RETURN 
 KIT DONE (MODIFIED)   SAVE USER KIT