Calibrating for Increased Measurement Accuracy
Minimizing Error When Using Adapters
9. Perform the desired calibration with this new user kit.
10.Connect the test device as shown in 
 and measure the device.
Minimizing Error When Using Adapters
To minimize the error introduced when you add an adapter to a measurement system, the 
adapter needs to have low SWR or mismatch, low loss, and high repeatability.
Figure 6-20 Adapter Considerations 
In a reflection measurement, the directivity of a system is a measure of the error 
introduced by an imperfect signal separation device. It typically includes any signal that is 
detected at the coupled port which has not been reflected by the test device. This 
directivity error will add with the true reflected signal from the device, causing an error in 
the measured data. Overall directivity is the limit to which a device’s return loss or 
reflection can be measured. Therefore, it is important to have good directivity to measure 
low reflection devices.
For example, a coupler has a 7 mm connector and 40 dB directivity, which is equivalent to 
a reflection coefficient of 
ρ
 = 0.01 directivity in dB = 
20 log (
ρ)
. Suppose we want to 
connect to a device with an SMA male connector. We need to adapt from 7 mm to SMA.
If we choose a precision 7 mm to SMA adapter with a SWR of 1.06, which has 
ρ
=0.03, the 
overall directivity becomes 
ρ
=0.04 or 28 dB. However, if we use two adapters to do the 
same job, the reflection from each adapter adds up to degrade the directivity to 17 dB. The 
last example, shown in 
uses three adapters that shows an even worse 
directivity of 14 dB. It is clear that a low SWR is desirable to avoid degrading the 
directivity of the system.