Honeywell HA40 Manual Do Utilizador

Página de 39
HA40 4-Channel Controller
Revision 1 (07/08)
HA40 Technical Manual
2-6
2�2�5 Cal Setup Menu
The  CAL  SETUP  feature  supports  pushbutton  calibration  of  zero 
and span values.  This feature should only be utilized when there are 
no other zero/span controls within the monitoring system since it is 
inappropriate to calibrate a signal at more than one point.  Therefore, 
if calibration will be performed at another transmitter or monitoring 
device, the CAL SETUP feature should not be used.
The 
CAL SETUP menu allows entering the correct Zero Gas & Span 
Gas  set-point  values  needed  to  calibrate  the  channel.   These  are 
entered in the same engineering units as input range.
Figure 2-5. Cal Setup Menu
2�2�6 Calibrate Input Menu
The CAL SETUP flow chart (shown in Figure 2-6) is designed to make 
calibration quick, easy and error free.  A successful ZERO and SPAN 
calibration requires only a few keystrokes.  Optional 4-20mA outputs 
(if equipped) transmit 1.5mA during calibration and 4mA during the 
subsequent CAL DELAY to prevent triggering external alarms during 
calibration.  Local HA40 alarm relays are inhibited during calibration.  
Unintentional calibrations may be reset using the Set UNITY menu 
item.  Set UNITY resets Cal OFFSET to 0 & Cal GAIN to 1 which is 
useful for returning the calibration to a known starting place.  Sensor 
aging may be monitored by recording zero and span readings at Unity 
Gain when the sensor is new, and again later when degradation may 
have  occurred.    CAL  MODE  automatically  exits  if  no  keystroke  is 
detected after 5 minutes.
Use the following step-by-step procedure to perform ZERO and SPAN 
calibrations.
To enter the CAL SETUP from any data display, press the dual 
1. 
purpose down arrow / CAL key then use the up/down arrow 
keys to select the channel to calibrate.
Stimulate  the  monitor  to  be  calibrated  with  an  appropriate 
2. 
ZERO calibration standard.  Observe the screen’s live reading 
and when it is stable press the Edit key to perform the ZERO 
calibration.
If the ZERO calibration is successful, CAL SETUP automatically 
3. 
proceeds to the SPAN check.
Apply the correct SPAN calibration standard.  After the reading 
4. 
is stable, press the Edit key to perform a SPAN calibration.
 Warning
The SPAN calibration standard used must match the value specified
since this is the reading the HA40 will indicate after a successful 
SPAN calibration.  The SPAN calibration value may be edited if it 
becomes necessary to apply a different calibration standard to set 
SPAN (see Span Calibration in section 2.2.5).
If  the  SPAN  calibration  is  successful,  the  display  flashes 
5. 
“REMOVE CAL GAS” and starts the CAL DELAY.
CAL SETUP is complete after the end of the CAL DELAY.
6. 
The  flow  chart  in  Figure  2-6  illustrates  the  above  procedure.    UP, 
CAL, Next & Edit labels indicate keystrokes (down arrow/CAL is a 
dual purpose key).  The CAL SETUP information screen (top of the 
chart) is available for advanced users to see Offset / Gain calibration 
constants and live analog to digital converter (A/D) counts.  Span set 
point calibration values may also be edited from this screen.   Holding 
the up key, for 5 seconds during CAL SETUP, displays this screen.