4 • Troubleshooting
Doc. 065291-01 3/09
91
Flow rate is set to 0
Set the flow rate from the ICS-2100 Control panel in Chromeleon or from the 
touch screen 
HOME
 page.
Eluent valve is closed
Open the eluent valve from the ICS-2100 Control panel in Chromeleon or 
from the touch screen 
PUMP
 page.
Pump not primed
Broken pump piston
Replace the piston (see 
Eluent generator degas tubing assembly ruptured
Replace the EluGen cartridge holder (see 
4.7
Erratic Flow/Pressure Reading
Pump needs priming
Figure 4-2.  Waste and Priming Valves
Waste Valve
Priming Valve