ICS-2100 Ion Chromatography System
92
Doc. 065291-01 3/09
Damaged piston seal
Replace the piston seal (see 
Dirty pump check valve
Clean or replace the check valve (see 
Leaking liquid lines or fittings
Check the liquid lines and fittings for small leaks. Tighten or, if necessary, 
replace the liquid line connection (see 
). Refer to Installation of 
Dionex Liquid Line Fittings (Document No. 031432) for tightening 
requirements. The manual is included on the Dionex Reference Library CD-
ROM (P/N 053891).
4.8
Excessive System Backpressure
Restriction in the system plumbing
Check all liquid lines for crimping or blockage. Make sure the ferrule fittings 
are not overtightened onto tubing. Refer to Installation of Dionex Liquid Line 
Fittings
 (Document No. 031432) for details. The manual is included on the 
Dionex Reference Library CD-ROM (P/N 053891). 
If you have trouble isolating the restriction, refer to 
Plugged or damaged fitting
Isolate the faulty fitting by loosening fittings, one by one, until the pressure 
returns to normal. Repair or replace the fitting (see 
Flow rate through the columns too high
Set the correct rate for your application. If an incorrect flow rate calibration is 
suspected, calibrate the flow (see 
Clogged column bed supports
Refer to the instructions in the column manual for troubleshooting guidance. 
Column manuals are included on the Dionex Reference Library CD-ROM 
(P/N 053891).
Contaminated columns
Clean the columns as instructed in the column manual or replace the guard 
column.