National Instruments 1000DC 用户手册

下载
页码 42
Chapter 2
Configuring and Installing the SCXI Chassis
2-10
ni.com
Figure 2-8.  SCXI-1000DC Chassis Address Jumper Settings
Changing the Chassis Address
While referring to Figures 2-2 and 2-6, complete the following steps to 
change the chassis address of the SCXI-1000DC:
1.
Power off the chassis and remove the power cord from the power entry 
module.
2.
To prevent a shock hazard, wait at least one minute before proceeding 
to step 3.
3.
Using a screwdriver, remove the four (six on some revisions) panhead 
screws from the front panel of Slot 0.
4.
Remove the six screws from the rear panel.
5.
Remove the controller from Slot 0.
6.
Set all three jumpers for the chassis address you want.
7.
Replace the controller in Slot 0.
8.
Replace the four (six on some revisions) screws. Do not overtighten.
9.
Replace the six screws in the rear panel. Do not overtighten.
W3
+16
W3
+16
W2
+8
W2
+8
0
2
4
6
W1
1
3
5
7
0
2
4
6
W1
1
3
5
7
Chassis Address 0
Factory Setting
Chassis Address 19
Example Setting