National Instruments 1000DC 用户手册

下载
页码 42
Chapter 2
Configuring and Installing the SCXI Chassis
© National Instruments Corporation
2-11
Selecting Voltage and Replacing the Fuse for the SCXI-1000 
and SCXI-1001
If you ordered the chassis with the appropriate part number (the -0x 
extension of the kit part number corresponds to your geographical region), 
the voltage tumbler and fuse are correct for operation in your geographical 
region. Check the voltage on the voltage tumbler to ensure that you have 
the correct voltage tumbler setting and fuse.
The SCXI chassis can operate with line voltages of 100, 120, 220, and 
240 VAC. The voltage tumbler in the power entry module determines the 
voltage for which the chassis is configured. You can identify the operating 
voltage by looking at the number on the power entry module when the door 
that covers the tumbler is closed. The fuse is 5
× 20 mm, which has a 
current rating relative to the operating voltage. Table 2-4 shows the proper 
voltage selections and fuse ratings for different regions.
Caution
For continued protection, replace fuses only with fuses of the same type and 
rating.
 
Table 2-4.  SCXI-1000/1001 Voltage Selection and Fuse Ratings by Region
SCXI 
Chassis 
Type
Region/Voltage
Main 
Power Fuse 
Quantity
Main Power
Fuse Type
Backplane 
Fuse 
Quantity
Backplane
Fuse Type
SCXI-1000
100 V Japan
1
3/4 A 250 V 5 mm 
× 
20 mm SLO-BLO
2
1.5 A 125 V LittleFuse 
R251 01.5 TI
120 V
1
1/2 A 250 V 5 mm 
× 
20 mm SLO-BLO
2
1.5 A 125 V LittleFuse 
R251 01.5 TI
220 V Swiss
1
1/4 A 250 V 5 mm 
× 
20 mm SLO-BLO
2
1.5 A 125 V LittleFuse 
R251 01.5 TI
240 V Australia
1
1/4 A 250 V 5 mm 
× 
20 mm SLO-BLO
2
1.5 A 125 V LittleFuse 
R251 01.5 TI
240 V Europe
1
1/4 A 250 V 5 mm 
× 
20 mm SLO-BLO
2
1.5 A 125 V LittleFuse 
R251 01.5 TI
240 V North American
1
1/4 A 250 V 5 mm 
× 
20 mm SLO-BLO
2
1.5 A 125 V LittleFuse 
R251 01.5 TI
240 V U.K.
1
1/4 A 250 V 5 mm 
× 
20 mm SLO-BLO
2
1.5 A 125 V LittleFuse 
R251 01.5 TI