Renishaw plc RMP60 用户手册

下载
页码 42
33
Fault-finding
Symptom
Cause
Action
Probe crash
Inspection probe using
Review program
tool-setting probe signals.
Review installation.
Probe length offset
Review probe software.
missing/incorrect
Workpiece obstructing probe
Review program.
path.
No LED’s lit on RMI
No power to RMI
Check  wiring
RMI status LED’s do not
Radio link failure – RMP60
Check position of RMI,
correspond to RMP60
out of RMI range.
see performance envelope.
status LED’s
RMP60 has been enclosed/
Review installation
shielded by metal.
RMP60 and RMI are not
Partner RMP60 and RMI.
partnered.
RMI probe status LED
Dead batteries.
Change batteries.
continually lit red
RMI error LED lit during
Damaged cable
Check wiring.
probing cycle
Loss of power
Check wiring.
Dead batteries
Change batteries.
Probe false trigger
Increase spring pressure.
Reduce acceleration.