Fault-finding
Symptom
Cause
Action
Poor measurement
Debris on part or stylus.
Clean.
results.
Recalibrate if probe was
calibrated with debris on
stylus.
Repeatability of probe
Verify by repeated toolchange
into spindle.
and single point move.
Loose probe to shank
Check and tighten as
mounting or stylus.
required, recalibrate.
Offsets not being updated
Review software.
Calibrated feature has moved.
Check.
Measurement occurs as
Review software.
stylus leaves surface.
Calibration and probing
Review software.
speeds different.