Agilent Technologies N5700 User Manual

Page of 128
Load Considerations 
28 
Series N5700 User’s Guide  
•  Turn off the power supply. 
•  Remove the local sense jumpers from the J2 connector. 
•  Connect the negative sense lead to terminal 5 (-S) and the 
positive sense lead to terminal 1 (+S). Make sure that the 
connector plug is securely inserted into the connector body. 
•  Turn on the power supply. 
NOTE 
If the power supply is operating in remote sense and either the positive or 
negative load wire is not connected, an internal protection circuit will activate 
and shut down the power supply. To resume operation, turn the power supply 
off, connect the open load wire, and turn on the power supply. 
Load Considerations 
Multiple Loads 
The following figure shows multiple loads connected to one power 
supply. Each load should be connected to the power supply’s output 
terminals using separate pairs of wires. It is recommended that each 
pair of wires will be as short as possible and twisted or shielded to 
minimize noise pick-up and radiation. The sense wires should be 
connected to the power supply output terminals or to the load with 
the most critical load regulation requirement. 
-
-
Rem.sense
Local sense
ocal sense
Rem.sense
+L
+
Load lines,  twisted pair, 
shortest length possible.
+V
-V
Load#1
+
Load#3
+
Load#2
+
Power 
Supply
-
-
Rem.sense
Local sense
ocal sense
Rem.sense
+L
+
Sense lines.
Twisted pair or
+V
-V
Load
+
Power
Supply
Load lines. Twisted pair
shortest length possible.
shielded wires.